RPI FIELD SERVICE
CALIBRATION KIT
RPI Part #MIK074
For servicing the Midmark Ritter Ultra Clave
®
CALL (800) 221-9723 or (818) 882-8611 • FAX (818) 882-7028 • E-MAIL [email protected] • WEBSITE www.rpiparts.com
For use with these models only:
M9 (S/N Series: CZ, DA, DB, DX, DY, FD & OM)
M9D (S/N Series: FF & FG)
M11 (S/N Series: ES, ET, FP, FR & GB)
M11D (S/N Series: GC & GC)
Przeglądanie stron 0
1 2 3 4 5 6 ... 15 16

Podsumowanie treści

Strona 1 - CALIBRATION KIT

An ISO 9001 Certified CompanyRPI FIELD SERVICECALIBRATION KITRPI Part #MIK074For servicing the Midmark • Ritter Ultra Clave®CALL (800) 221-9723 or (81

Strona 2 - TABLE OF CONTENTS

9AT-A-GLANCECALIBRATION PROCEDURE1) Open door.2) Depress LIQUIDSand PACKS buttons at the same time as plugging in thepower cord.3.) Starting with UNWR

Strona 3 - 1.0 - LEAK DETECTION GUIDE

10Figure 1 – Diagnostics Procedure (M9 & M11 and M9D & M11D Style Key Pad Displays)M9 & M11 Style Key Pad DisplayM9D & M11D Style Key

Strona 4

5.0 - CALIBRATION PROCEDURE - PRESSURE/TEMPERATURE POTENTIOMETER ADJUSTMENTS (continued)11NOTES: 1) Turning the adjusting screw of the pressure range

Strona 5

8) CONNECTING THE TEST PRESSURE GAUGE TO THE STERILIZER: Disconnect the black tubing from Pressure Sensor and from the barb fitting.Keep black tubing

Strona 6 - ERROR CODE 7 ERROR CODE 10

13Figure 2 – Old Style PC BoardNOTE: The pressure zero, pressure range, and temperature potentiometers must not be adjusted exclusively. If one potent

Strona 7

14NOTE: The pressure zero, pressure range, and temperature potentiometers must not be adjusted exclusively. If one potentiometer needs to be adjusted,

Strona 8

RPI PART # OEM PART # DESCRIPTIONMIP050 002-0357-00 TEMPERATURE PROBE ASSEMBLY • •MIS075 002-0358-00 WATER LEVEL SENSOR ASSY • •PCV052 002-0359-00 PRE

Strona 9

This guide has been prepared to help troubleshoot, evaluate and repair the Midmark•Ritter M9 and M11 Ultra-Clave®.TABLE OF CONTENTS1.0 – Leak Detectio

Strona 10 - (continued)

21.0 - LEAK DETECTION GUIDEDOOR and/or DAM GASKET Check for water or steam leaking around the door.VENT SOLENOID VALVE Check for water or steam leakin

Strona 11

FILTERS: MIF038 or MIF062Depending on the age of the machine, you will need to replaceeither the sintered bronze style Chamber Filter (MIF038) or both

Strona 12

4Power interruption to control PC board.1. Overheat thermostat activates (opens).• Pressure leak during sterilization cycle.• Water sensor shorted (wh

Strona 13

5Hardware and/or software problem – Watchdog timer reset error.1. Poor electrical connection or faulty wiring (bad power source tosterilizer).2. Main

Strona 14 - Figure 2 – Old Style PC Board

6Printer FuseFast Acting; 1/4" dia. x 1-1/4" Lg.; 3 Amp, 250 VoltLine Fuse(s)Fast Acting; 1/4" dia. x 1-1/4" Lg.; 15 Amp, 250 Volt

Strona 15 - Figure 3 – New Style PC Board

74.0 - DOOR OPENING TEST PROCEDURENote: This test will only verify that the door opening relay on the main PC board is functioning. A completecycle is

Strona 16 - M9 & M11 UltraClave

8IMPORTANT The Calibration Procedure that begins onpage 9 requires the sterilizer to be plugged in andpowered up while the adjustments are being per-