Printer Fuse
Fast Acting; 1/4" dia. x 1-1/4" Lg.; 3 Amp, 250 Volt
Line Fuse(s)
Fast Acting; 1/4" dia. x 1-1/4" Lg.; 15 Amp, 250 Volt
If sterilizer is overheating in Dry mode, check the settings.
Caution: Disconnect the power cord before changing these settings.
Off = Position switch to the left. On = Position switch to the right.
Ground Circuit continuity failed the (5) minute time limit.
If water is continuously filling the chamber:
Check for an open circuit or corrosion on the Water Lever Sensor.
Check temperature probe by disconnecting it from Main PC board.
The fill valve may be stuck in the Open position.
If water flow is restricted or not present:
Check for debris (dirt) on the filter and in the fill lines.
Check fill valve for proper operation or debris.
Water level sensor malfunction.
Sterilizer displays temperature and pressure, but the LED status
lights do not light up.
Disconnect the power cord and wait 30 seconds (this will reset the
internal CPU). Run additional cycles.
DRYING CYCLE DIP SWITCH SETTING
FUSE CHART (Fuses located on Main PC Board) LOW WATER ERROR
DISPLAY LIGHTS INACTIVE
3.0 - ERROR CODES AND OTHER DIAGNOSTICS (continued)
Coolest: Setting #1
#1 OFF
#2 OFF
#3 ON
#4 ON
Cool: Setting #2
#1 OFF
#2 OFF
#3 OFF
#4 ON
Warm: Setting #3
#1 OFF
#2 OFF
#3 OFF
#4 OFF
Warmest: Setting #4
#1 OFF
#2 OFF
#3 ON
#4 OFF
If sterilizer needs to
dry even cooler -
Setting #5
#1 OFF
#2 ON
#3 ON
#4 ON