2
1.0 - LEAK DETECTION GUIDE
DOOR and/or DAM GASKET Check for water or steam leaking around the door.
VENT SOLENOID VALVE Check for water or steam leaking back into the reservoir through the condensing coil.
FILL SOLENOID VALVE Check for water leaking back into the reservoir through the fill line by observing an increase in the reservoir
water level and/or presence of bubbles.
BELLOWS VALVE Check for excessive steam from the bellows line in the reservoir. Note: a slight "hiss" of steam is not
uncommon.
PRESSURE RELIEF Check for water leaking beneath the back of the sterilizer.
VALVE Caution: Release pressure in the sterilizer before replacing the pressure relief valve.
PRESSURE Check for steam leaking onto or around the Main PC board.
SENSOR HOSE