4
Power interruption to control PC board.
1. Overheat thermostat activates (opens).
• Pressure leak during sterilization cycle.
• Water sensor shorted (which will result in bypassing of filling
phase).
• Chamber water level too low.
• High input voltage during drying cycle.
• Faulty overheat thermostat.
2. Poor electrical connection or faulty wiring (bad power source to
sterilizer).
3. Main PC board may be faulty.
3.0 - ERROR CODES AND OTHER DIAGNOSTICS
The Stop button was depressed during the sterilization cycle.
The On/Standby
button was depressed during the sterilization cycle.
Excessive temperature (Temperature exceeds 277˚F).
• Check for pressure leaks. Refer to 1.0 - Leak Detection Guide.
• Temperature probe is dirty or malfunctioning.
• Sterilizer may need to be re-calibrated. Refer to 5.0 - Calibration
Procedure.
Excessive pressure (Pressure exceeds 35 psi).
• Sterilizer is overloaded.
• Bellows is closing prematurely trapping air in the chamber.
• Temperature probe is dirty or malfunctioning.
• Sterilizer may need to be re-calibrated. Refer to 5.0 - Calibration
Procedure.
Door ajar during sterilization cycle.
• Verify the door latch properly engages the door switch and the switch
activates.
• Check door switch continuity and the wiring connections.
ERROR CODE 1
ERROR CODE 2
ERROR CODE 5
ERROR CODE 6
ERROR CODE 3
ERROR CODE 4