Power interruption to control PC board.
1. Overheat thermostat activates (opens).
Pressure leak during sterilization cycle.
Water sensor shorted (which will result in bypassing of filling
phase).
Chamber water level too low.
High input voltage during drying cycle.
Faulty overheat thermostat.
2. Poor electrical connection or faulty wiring (bad power source to
sterilizer).
3. Main PC board may be faulty.
3.0 - ERROR CODES AND OTHER DIAGNOSTICS
The Stop button was depressed during the sterilization cycle.
The On/Standby
button was depressed during the sterilization cycle.
Excessive temperature (Temperature exceeds 277˚F).
Check for pressure leaks. Refer to 1.0 - Leak Detection Guide.
Temperature probe is dirty or malfunctioning.
Sterilizer may need to be re-calibrated. Refer to 5.0 - Calibration
Procedure.
Excessive pressure (Pressure exceeds 35 psi).
Sterilizer is overloaded.
Bellows is closing prematurely trapping air in the chamber.
• Temperature probe is dirty or malfunctioning.
Sterilizer may need to be re-calibrated. Refer to 5.0 - Calibration
Procedure.
Door ajar during sterilization cycle.
Verify the door latch properly engages the door switch and the switch
activates.
Check door switch continuity and the wiring connections.
ERROR CODE 1
ERROR CODE 2
ERROR CODE 5
ERROR CODE 6
ERROR CODE 3
ERROR CODE 4